- + 91 265 2632148 / 49
- haresh@lablineindia.com
Product Range
Advance Digital Centrifuge
- Brushless
Digital Centrifuge - Small Foot print
Brushless centrifuge - Swing
60 - Advance
Digital Centrifuge - Digital
Centrifuge - Hematocrit
Centrifuge - Medium
Centrifuge - Small
Centrifuge
-
Capacity :As per Selected RotorMain input :220v , 50Hz / 110v , 60HzMax. Speed :As per Selected RotorSize :(W) 355 x (D) 410 x (H) 280 mmWeight :14 Kg. NetSpeed Regulator :Programmable Preset RPM With
+ / -200 RPMRPM Display :LCD type preset & actaul RPMTimer Regulator :Programmable Preset TimeTimer Display :LCD type preset & remaining time up to 30 minutesRotors
RT-01
RT-02
RT-03
RT-04
RT-05
RT-06
RT-07
RT-08
RHC-12
RHC-24
MICRO TUBE
RD-11
Rotor's Range
Model No Rotor Model Rotor Capacity Type of Rotor Min RPM Max RPM Max RCF CF-708 RT-01 8 TUBE x 15 ML SWING OUT 1500 3500 1950 CF-700 RT-02 12 TUBE x 5 ML SWING OUT 1500 4000 2000 CF-711 RT-03 8 TUBE x 15 ML ANGULAR 1500 5000 3100 CF-722 RT-04 12 TUBE x 15 ML ANGULAR 1500 5000 3650 CF-733 RT-05 16 TUBE x 15 ML ANGULAR 1500 4500 3350 CF-744 RT-06 24 TUBE x 5 ML ANGULAR 1500 5000 2900 CF-755 RT-07 16 TUBE x 15 ML SWING OUT 1500 3500 1900 CF-766 RT-08 24 TUBE x 5 ML SWING OUT 1500 4000 1950 CF-777 RHC-12 12 Capillaries of
QBC - ll Plus SystemRHC-12 4000 12000 15300 RHC-24 24 Capillaries of size
75 mm x 1 mm boreRHC-24 4000 12000 15300 MICRO
TUBE24 TUBE x 1.5 ML / 2.0 ML ANGULAR 4000 12000 13500 Features
Motor :Universal (AC/DC Carbon Brush Type)Controller :It is microprocessor controlled and programmable
For pre set RPM and pre defined timeDisplay :It is LCD - 2 line display with back light it display
pre set RPM , actual RPM, pre defined TIME and
remaining TIMESafty :A. The centrifuge will not run if door is open.
B. The lid will lock immediately if rotor start spinning
C. The lid remain locked till rotor become standstill
D. If power supply fail then lid unlock immediately
but do not open lid for few minutes.Imbalance :The rotation stops if rotor is imbalance and imbalance
indicator LED - “ON” till restart of deviceBrake :The Device is Having automatic Brake with LED indicator.Thermal Protection :The motor is thermally protected